Télécharger Imprimer la page

Hettich EBA 200 Mode D'emploi page 69

Masquer les pouces Voir aussi pour EBA 200:

Publicité

Les langues disponibles

Les langues disponibles

7.1.2
Runtime t
7.1.3
Speed, RPM
7.1.4
Relative centrifugal force, RCF
AB1800en
[t] buttons to set the desired value.
7.
Use the
A numerical value from 10 mm to 250 mm can be set.
Adjustable in 1 millimetre increments
[SELECT] button.
8.
Press the
�� Relative centrifugal force
[t] buttons to set the desired value.
9.
Use the
A numerical value can be set that gives a speed between 200 RPM
and the maximum rotor speed.
Adjustable in 1 second increments.
[SELECT] button.
10.
Press the
�� The DEC brake level is displayed.
fast: short ramp-down time
slow: long ramp-down time
[t] buttons to set the desired value.
11.
Use the
12.
Press the
[START/PULSE] button.
�� The settings are stored.
[t] buttons to set the desired value
1.
Use the
�� The value is set up to 1 minute in 1 second increments.
The value is set from 1 minute in 1 minute increments.
Adjustable from 1 to 99 minutes and 1 to 59 seconds.
2.
The parameters t/min and t/sec must be set to zero to set continuous
operation.
'--:--' is displayed.
��
[RCF] button to select the RPM indicator.
1.
Press the
[RCF] button to toggle between the two parameters
�� Press the
'RPM' ) and RCF ( '>RCF<' ).
RPM (
[RPM/RCF] buttons to set the desired value.
2.
Use the
A numerical value from 200 RPM to the maximum rotor speed can be
set.
Adjustable in 10 second increments.
The relative centrifugal force RCF is dependent on the speed and the centri-
fuging radius.
The relative centrifugal force RCF is stated as a multiple of the acceleration
due to gravity (g).
The relative centrifugal force RCF is a dimensionless numerical value and is
used to compare the separation and sedimentation performance.
Rev.: 09 / 01.2024
Software operation
'RCF' is displayed.
25 / 39

Publicité

loading

Ce manuel est également adapté pour:

Eba 200 s