Operation
Fig. 15: Rotor installation and
removal
1
Marker bar
2
Motor shaft
3
Surfaces
Installing the rotor of the EBA
200
6.3
Loading
Filling centrifuge tubes
20 / 39
1.
Open the lid.
2.
Loosen the rotor clamping nut using the spanner supplied by turning
it anticlockwise.
1
�� After passing the working point for lifting the rotor, the rotor
2
detaches from the cone of the motor shaft (
3.
Turn the clamping nut until the rotor can be lifted off the motor shaft.
3
4.
Remove the rotor.
Personnel:
Trained user
The lid is open.
1.
Clean the motor shaft (
2.
Lightly grease the motor shaft (
cleaning and disinfection' on page 32 .
3.
Place the rotor vertically on the motor shaft (
The two marker bars (
3  ) on the motor shaft.
surfaces (
4.
Tighten the rotor clamping nut hand-tight by turning it clockwise using
the hex key provided.
5.
Check that the rotor is firmly seated.
6.
A test run must be performed if a different rotor has been installed.
For the test run, the adjustment weight supplied (7g) must be placed
in a rotor location and a centrifugation run with a runtime of 1 minute
must be performed at a speed of 6000 RPM.
�� The drive must not switch off.
WARNING
Risk of injury from contaminated sample material.
Contaminated sample material escapes from the sample
tube during centrifugation.
Use centrifuge tubes with special screw caps for haz-
ardous substances.
For risk group 3 and 4 materials, use a biosafety system
in addition to the sealable centrifuge tubes (see WHO's
'Laboratory Biosafety Manual').
Rev.: 09 / 01.2024
2  ) and rotor hole.
2 ), see ⮫  Chapter 8.2 'Instructions for
1  ) on the rotor must be parallel to the two
The adjustment weight must be removed from the
rotor location again before the next centrifugation
run.
2  ).
2  ).
AB1800en