Loading the angle rotors
AB1800en
NOTICE
Damage to the device due to highly corrosive substances.
Highly corrosive substances may impair the mechanical
strength of rotors, buckets and accessories.
−
Do not centrifuge highly corrosive substances.
Standard glass centrifuge tubes can be loaded up to RCF
4000 (DIN 58970 part 2).
Personnel:
■
Trained user
Fill centrifuge tubes outside the centrifuge.
The maximum capacity of the centrifuge tubes specified by the manu-
facturer must not be exceeded.
With angle rotors, the centrifuge tubes must only be filled to the
extent that no liquid can be ejected from the tubes during the centrifu-
gation run.
It must be ensured that there is a uniform fill level in the tubes in
order to keep the weight differences in the centrifuge tubes as low as
possible.
Personnel:
■
Trained user
1.
Check that the rotor is firmly seated.
2.
The centrifuge tubes must be distributed evenly over all locations on
the rotor.
No liquid must be allowed to enter the rotor and the centrifuging
chamber when loading the rotor.
With rotors, the centrifuge tubes must only be filled to the extent that
no liquid can be ejected from the tubes during the centrifugation run.
The weight of the permissible filling capacity is indicated on each
rotor. The weight must not be exceeded.
Rev.: 09 / 01.2024
Operation
21 / 39