Fig. 14: Transport lock
1
Spacer sleeve
2
Screw
5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
AB1800en
The lid is closed.
The mains cable is disconnected from the device.
1.
Tilt the device on the back of the device.
2.
Unscrew 2 screws (
3.
Remove 2 spacer sleeves (
4.
Keep the screws and spacer sleeves in a safe place.
1
2
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
NOTICE
Damage to the samples and the device if the ambient
temperature exceeds or falls below the respective max-
imum/minimum permissible ambient temperature.
Personnel:
Trained user
1.
Place the device on a stable and level surface.
2.
Maintain a distance of 300 mm around the device.
3.
Comply with the ambient conditions in the technical data (⮫ 
3.1 'Technical data' on page 9 ).
Rev.: 09 / 01.2024
2  ).
1  ).
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
Place the device on a stable and level surface.
Select the installation surface dependent on the weight
of the device.
Comply with the maximum and minimum permissible
ambient temperatures for installation of the device.
Do not place the device next to a heat source.
Do not expose the device to direct sunlight.
Do not expose the device to frost.