Télécharger Imprimer la page

Dräger Polytron 8000 Serie Notice D'utilisation page 49

Masquer les pouces Voir aussi pour Polytron 8000 Serie:

Publicité

Les langues disponibles

Les langues disponibles

Storage
Pressure
Humidity
Temperature
Time
Influence of environmental parameters
Zero-point
Temperature
-20 to 40 °C
≤ 0.1 Vol% O
-40 to 65 °C
≤ 0.2 Vol% O
Pressure
< ± 0.1 Vol% O
Humidity
≤ 0.1 Vol% O
When measuring oxygen in the presence of helium, helium
compensation can be used. The option of helium
compensation is no certified measuring function according to
EN 50104.
CAUTION
Sensor failure
The exposure of the sensor to higher concentrations of
unsaturated hydrocarbons, alcohols, or hydrogen over an
extended period of time (dose approx. 100,000 ppm x hours)
can cause the instrument to fail.
► Ensure that the sensor is not exposed to such
concentrations over an extended period of time.
Response time
t
0...20
0-5 Vol% O
2
≤ 14 seconds
0-25 Vol% O
≤ 10 seconds
2
1) Response time is higher than the permissible limit of EN 50104.
Stabilization time: 5 x t
0...90
Measuring range
0-5...25 Vol% O
, 0-25 Vol% O
2
Default value: 25 Vol% O
2
Minimum reading: -1.25 Vol% O
Sensor warm-up time
Operation: ≤ 20 minutes / Adjustment: ≤ 6 hours
Instructions for use
|
Dräger Polytron
no influence
30 to 70 % RH, non-condensing
(only relevant if sensor packag-
ing is open)
0 to +40 °C
Storage of sensors is not
intended. Sensors should be
used immediately. Remaining
life span = expected life span -
storage time
Sensitivity
≤ 0.1 Vol% O
2
2
≤ 0.5 Vol% O
2
2
≤ 0.8 Vol% O
2
2
≤ 0.9 Vol% O
2
2
t
0...90
≤ 45 seconds
1)
≤ 27 seconds
2
2
®
8000 Series SW version ≥ 4.0.x
18.3.3 Dräger SensorAlive (6809630)
The Dräger SensorAlive (6814014) is an electrochemical
diffusion sensor with an integrated gas generator for
measuring hydrogen sulfide (H
Environmental parameters
Pressure
Humidity
Temperature
Storage
Pressure
Humidity
Temperature
Time
Influence of environmental parameters
Pressure
Humidity (at +40 °C)
Temperature
Interference gases (e.g., NH
the sensor.
Response times in diffusion operation
t(50)
t(90)
Decay times in diffusion operation
t(50)
t(10)
Performance Approvals
S) in ambient air.
2
20.7 to 35.5 in. Hg (700 to
1200 hPa)
5 to 95 % RH, non-condensing
-40 to +65 °C
no influence
30 to 70 % RH, non-condensing
(only relevant if sensor packag-
ing is open)
0 to +40 °C
Storage of sensors is not
intended. Sensors should be
used immediately. Remaining
life span = expected life span -
storage time
Relative deviation from the
reading
At 1000 hPa: ≤ 0.15 ppm
at 0 and 95 %RH: ≤ 0.29 ppm
at 20 °C: ≤ 0.44 ppm
) can impair the function of
3
≤ 20 s
≤ 55 s
≤ 20 s
≤ 55 s
|
en
49

Publicité

loading