7.5
Menus
7.5.1 Navigating in a menu
Button Function
Scrolls upwards.
Sets values.
Scrolls downwards.
Sets values.
Confirms entry.
OK
Selects menus and functions.
7.5.2 Opening menus
● To open the Information menu:
a. Tap and hold
for 1 second in measuring mode.
● To open the Calibration menu:
a. Tap and hold
for 3 seconds in measuring mode.
b. Select Enter password.
c. Enter the corresponding password and confirm with
[OK].
● To access all menus:
a. Tap and hold
for 3 seconds in measuring mode.
b. Select Enter password.
c. Enter the corresponding password and confirm with
[OK].
7.5.3 Passwords
Calibration
Permits access to the Information menu
and the Calibration menu (adjustment
PWD
settings)
Default value: _ _ _ 1
Permits access to all configurations and
Settings PWD
menus
Default value: _ _ _ 2
Information on changing the passwords can be found in 11.1.
7.5.4 The Information menu
The Information menu can be used to access the following
information:
Instrument
Warnings
Displays warning messages in plain text
and the corresponding number.
If there are several warning messages
available, there is an indicator (e.g. 1/3 =
Screen 1 of 3).
Displays error messages in plain text and
Faults
the corresponding number.
If there are several faults, there is an indica-
tor (e.g. 1/3 = Screen 1 of 3).
Device flag This function displays warning and fault
codes as a table. If all codes are 00 there
are no warning or fault messages available.
Instructions for use
|
Dräger Polytron
®
8000 Series SW version ≥ 4.0.x
Displays an overview of the installed hard-
Modules
ware modules and the software version of
the EC sensor interface named SIOS. To
access detailed information select the
appropriate module.
= installed modules
= modules not installed
Sensor
The calibration date, unit, gas and concen-
Last cal.
date
tration of the last calibration are displayed.
Next cal.
Displays the next calibration due date.
date
The sensor vitality is displayed in %. Dräger
1)
Vitality
recommends replacing the sensor if the
vitality is < 25 %.
Sensor
The current and the highest sensor tem-
perature is displayed.
tempera-
1)
ture
Sen-
The sub-menu shows the days until the end
sorAlive
of the service life and the vitality in %.
Data logger
Logger
Displays the data logger status (active or
non-active)
status
Displays the history of the past 15 minutes
Graph
on a time/concentration graph.
1) Function only available with diagnostic dongle.
Operation
|
en
19