Treatment READY mode

• Touching the
STANDBY/READY
STANDBY mode switches the system, after about three seconds,
into Treatment READY mode (Fig. 25).
1
Laser warning lamp
Fig. 25

Treatment READY mode

In Treatment READY mode, the aiming beam and the laser warning
lamp are always on.
Parameters are set in the same way as for Treatment STANDBY
mode (see page 42).
When the firing switch is activated, the laser beam is fired according to
the parameters set.
Warning
Never fire the laser if the aiming beam is not clearly visible in the target
area.
• Touching the
STANDBY/READY
into Treatment STANDBY mode.
Note
If, in Treatment READY mode, more than five minutes
elapses without the laser being fired, the system returns to
Treatment STANDBY mode.
000000-1354-664 VISULAS YAG III 15.08.2005
Operation
button whilst in Treatment
button switches the system back
1
47