In treatment mode, the therapy beam can be applied via the
LSL YAG III Combi laser slit lamp.
All treatment-related laser parameters are displayed on the control
panel and can be changed in this mode.
• To set individual laser parameters, touch the appropriate parameter
button and select the value required by turning the control knob.
• Select the pulse mode required.
• Pressing the control knob takes you directly to the entry field for the
laser energy of the therapy beam. Rotate the knob to set the energy
required.
• The aiming beam can be turned on and off using the
BEAM
button. When the aiming beam is on, this button is
highlighted.
• After touching the
BRIGHTNESS
beam can be adjusted by rotating the control knob.
• A pulse counter is also displayed. This can be reset to zero by
touching the
RESET
• Touching the
ARROW
• Touching the button for switching between treatment modes
STANDBY/READY
into Treatment READY mode. This transition time is visually
indicated by a blinking button.
000000-1354-664 VISULAS YAG III 15.08.2005
button, the intensity of the aiming
button.
button returns you to the Main menu.
switches the system, after about three seconds,
Operation
AIMING
53