Perform online update/upgrade:
Before you perform online update/upgrade, ensure that
online updates/upgrades have been enabled (see below) and
a Wi-Fi network connection has been set up (see section 6.1).
Check that the network you connect to can reach the address:
https://api.services.ambu.com to ensure that the displaying
unit can retrieve any available update/upgrade.
•
Press the Settings tab, then press About.
•
Press System upgrade, then press Online upgrade
•
Wait while the displaying unit is checking for available updates/upgrades.
•
If a new software version is available, press Upgrade software
•
Follow the instructions on the screen to finish installing the update/upgrade.
Note: Software updates/upgrades cannot be carried out when an endoscope is connected to
the displaying unit or simultaneously with other functions or procedures.
13.3. Reporting a problem
If you are experiencing problems with the displaying unit, please follow the Troubleshooting
guide in chapter 13 to find a solution. If this does not solve the problem, please contact your
local Ambu representative. To solve the problem, Ambu may need a log file providing
information about the displaying unit system.
Export a log file:
•
Press the Settings tab, then press About.
•
Press Export log files.
•
Press the USB flash drive name,
then press the Export icon
•
Wait while the log files are exported,
then press Ok.
13.4. Data reset
If a displaying unit is to be handled by a third party or needs to be disposed, it is advised to
delete all sensitive data. Please follow your local guidelines for data protection. Only the
Administrator can reset the data.
The data reset function will delete:
•
All procedure folders with saved images and videos including patient information
retrieved from the Worklist.
•
All created user profiles, the Administrator and Service user profiles will be reset to
default password.
•
Wireless network settings and certificates.
•
DICOM, PACS and Worklist server settings
and certificates.
Reset the data:
•
Press the Settings tab, then press About.
•
Press Data reset, then press Delete
A dialog box appears informing of the
consequences and asking to confirm.
•
Press Confirm, then enter Administrator
password and press Continue.
•
A dialog box appears informing that this
operation may take a while, the operation can succeed or fail. If failed, please try again.
14. Cleaning and Disinfection of the Displaying Unit
The displaying unit is a reusable medical device. According to the Spaulding classification,