Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity test
Voltage
Dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted
Radio
Frequency
IEC 61000-4-6
Radiated Radio
Frequency
IEC 61000-4-3
44
IEC 60601-1-2
test level
0 % U
for 0,5 cycle
T
0 % U
for 1 cycle
T
70 % U
for
T
25 cycles
0 % U
for
T
250 cycles
30 A/m
3 V RMS
0,15 MHz – 80 MHz
6 V RMS
in ISM bands
80 % AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
27 V/m
385 MHz
PM, 18 Hz, 50 %
28 V/m
450 MHz
PM, 18 Hz, 50 %
9 V/m
710 MHz
PM, 217 Hz, 50 %
9 V/m
745 MHz
PM, 217 Hz, 50 %
9 V/m
780 MHz
PM, 217 Hz, 50 %
28 V/m
810 MHz
PM, 18 Hz, 50 %
Compliance Level
0 % U
for 0,5 cycle
T
0 % U
for 1 cycle
T
70 % U
for
T
25 cycles
0 % U
for
T
250 cycles
30 A/m
3 V RMS
0,15 MHz – 80 MHz
6 V RMS
in ISM bands
80 % AM at 1 kHz
3 V/m
80 – 2700 MHz
80 % AM at 1 kHz
27 V/m
385 MHz
PM, 18 Hz, 50 %
28 V/m
450 MHz
PM, 18 Hz, 50 %
9 V/m
710 MHz
PM, 217 Hz, 50 %
9 V/m
745 MHz
PM, 217 Hz, 50 %
9 V/m
780 MHz
PM, 217 Hz, 50 %
28 V/m
810 MHz
PM, 18 Hz, 50 %
Electromagnetic
Environment Guidance
Mains power quality shall
be that of a typical
commercial or hospital
environment.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Portable and mobile
RF communications
equipment should be
used no closer to any part
of the system, including
its cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of
the transmitter.
Field strengths from
fixed RF transmitters,
as determined by
an electromagnetic
site survey:
a) Should be less than the
compliance level in each
frequency range.
b) Interference may
occur in the vicinity of
equipment marked with
the following symbol.