Problem
The displaying
unit does not
turn ON when the
power button
is pressed.
No live image
from the
visualization
device on the
displaying unit
screen
OR
No LED light in
the distal end of
the visualization
device.
Live image on the
displaying unit is
ONLY working
while booting.
40
Possible cause
The displaying unit is not
connected to power.
Ensure that the light indicator
in the power button lights up
when the displaying unit is
connected to power.
The displaying unit does not
turn ON even though the power
is connected.
A recorded image in the current
procedure folder is blocking the
live image
OR
A menu box is blocking
the live image.
LED light switched off
by activation of the Light
off function.
Failure on visualization
device connection.
The displaying unit and the
Ambu visualization device have
communication problems.
Failure on visualization
device or VDI-port on
the displaying unit.
Processing failure.
Recommended action
Connect the power cable to the
displaying unit and to a power
outlet. Make sure the power
outlet is turned on.
Perform a Power OFF Cycle
(follow the procedure above this
table).
Return to live image by pressing
the Live View tab
If this does not work perform a
Power OFF Cycle (follow the
procedure above this table).
Deactivate the Light off function
(see section 7.3.8.).
Connect/reconnect visualization
device.
Perform a Power OFF Cycle
(follow the procedure above this
table).
Connect a new
visualization device.
If this does not work, the VDI-
port on the displaying unit may
be damaged.
Perform a Power OFF Cycle
(follow the procedure above this
table).
.