In this example, the instrument is in Power Supply mode.
– Set Voltage Mode to List: Press V Mode Lst.
– Set Current Mode to List: Press Curr Mode Lst.
– Set Trigger Source to Key. Press Trig Src Key.
– Press the color-coded [On] key to enable the output channel.
– Press Run to start the list operation. To abort the operation, press Stop.
– Press Back to exit and return to the previous menu.
From the remote interface:
To set the sequencer to List mode:
TRAN:MODE LIST, (@1)
To configure the List voltage with 1 V, 2 V, 3 V, 4 V and 5 V:
LIST:VOLT 1,2,3,4,5, (@1)
To configure the List current with 0.1 A, 0.2 V, 0.3 A, 0.4 A and 0.5 A:
LIST:CURR 0.1,0.2,0.3,0.4,0.5, (@1)
To configure all the List time with 1 s:
LIST:DWELL 1,1,1,1,1, (@1)
To enable all the List BOST:
LIST:TOUT:BOST 1,1,1,1,1, (@1)
To disable all the List EOST:
LIST:TOUT:EOST 0,0,0,0,0, (@1)
To set the voltage mode to List:
VOLT:MODE LIST, (@1)
To set the current mode to List:
CURR:MODE LIST, (@1)
To set the trigger source to Key/Immediate:
TRIG:SOUR IMM, (@1)
To enable the output channel:
OUTP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)