Fields
Voltage
Sets the voltage, current, resistance, or power value for selected step in voltage, ampere, ohms, or watts.
Current
Resistance
Power
Time
Sets the run-time for selected step in seconds.
BOST
Enables check box to set which step will generate a trigger-out signal at the beginning of the step (BOST).
EOST
Enables check box to set which step will generate a trigger-out signal at the end of the step (EOST).
For additional settings, press Properties to open the Sequencer List Properties window. Configure the Sequencer
List accordingly. Refer to the below table for details.
Power Supply mode
Settings
Available Key Set-
tings
Voltage/Current After
DC or List
List
Current After List
Voltage After List
Resistance After List
Power After List
Pace
Dwl or Trg
Keysight E36731A User's Guide
Description
Specifies what happens when the output sequence completes.
Default is DC.
Action required:
Press V/I List, Curr List, Volt List, Res List, or Pow List to toggle between DC and List.
DC (Return to DC
Returns to the DC value that was in effect before the output
Value)
sequence started.
List (Last List Value) Remains at the last list value.
Configures the pacing of the step.
Default is Dwl.
Action required:
Press Pace to toggle between Dwl and Trg.
Dwl(Dwell)
The next step immediately outputs when the dwell time has
elapsed.
Trg (Trigger)
The next step immediately outputs when an external trigger is
received. If the step time completes before trigger occurs, the step
remains at the last list value while waiting for the trigger.
Load mode
Description
95