Télécharger Imprimer la page

Keysight E36731A Manuel D'instructions page 114

Publicité

Les langues disponibles

Les langues disponibles

In Load mode:
Trigger Settings >
114
Trigger
V<1-2> Level,
Source
I<1-2> Level, Run
Key, List Key,
On/Off, DIO or
Remote
Level
Slope
Trigger Positions Specifies a trigger offset. This allows the specified percent of pre-
Configures the trigger source.
Default is Run Key.
Action required:
Press Trigger Source to select a trigger source. This trigger
source will trigger the data logging on all input channels.
V1 Level
Selects the voltage trigger level as
trigger source.
Triggers the measurement when the
voltage of the corresponding input
passes through the specified level.
I1 Level
Selects the current trigger level as
trigger source.
Triggers the measurement when the
current of the corresponding input
passes through the specified level.
Run Key (Run/Stop Key) Selects the Datalog Run Stop key as
a trigger source.
List Key (List Run/Stop
Selects the List Run/Stop key as a
Key)
trigger source.
On/Off (Input On/Off
Selects the Input On/Off keys as a
Key)
trigger source. Also applies to the All
Inputs On/Off key.
DIO (DIO Trigger In)
Selects any configured digital IO
(with Trigger Input function) as a
trigger source.
Remote (Remote
Selects a remote interface command
Command)
as a trigger source.
Specifies a trigger level if you selected a Voltage level or Current
level as the trigger source. Along with the level, you must also
specify a slope.
Specifies if the measurement will be triggered on the positive
(up-slope) or negative (downslope) portion of the waveform.
trigger data to be logged to the file. The trigger position is
expressed as a percentage of the data log duration. For example,
if you specified a data log duration of 30 minutes and a trigger
position of 50%, the Data Logger will log 15 minutes of pre-trig-
ger data to the file before the trigger occurs. Subsequently, 15
minutes of post-trigger data will then be logged to the data file.