Parameter
Available Key Set-
tings
Transient Setting
minimum to maximum Sets the transient voltage, current, resistance, or power level in voltage, ampere, ohms, or
(I
, V
, P
, or R
)
1
1
1
1
Current Mode
Fix, Stp, or Lst
Voltage Mode
Resistance Mode
Power Mode
Trigger Source
Key, IO, or Rmt
Trigger Delay
0 to 0.255 s
Step 2 – Run the input sequence
In this example, the load operation is in CC mode.
– Set I
, trigger delay and repeat count accordingly: Use the navigation keys to select the field; use the numeric
1
entry keys to enter the value. The value is set when you press [Enter].
– Set Trigger Source to Key. Press Trig Src Key.
– Press the color-coded [On] key to enable the input.
– Press Run to start the sequence operation. To abort the operation, press Stop.
– Press Back to exit and return to the previous menu.
Keysight E36731A User's Guide
watts.
Sets the current/voltage/resistance/power mode. This determines what happens to the input
current/voltage/resistance/power when the system is initiated or triggered.
Default is Fix.
Action required:
Press Curr Mode, Volt Mode, Res Mode, or Pow Mode to toggle between Fix, Stp, and Lst.
Fix (Fixed)
Keeps the input at its immediate value.
Stp (Step)
Steps the input to the triggered level when a trigger occurs
Lst (List)
Causes the input to follow the list values when a trigger occurs.
Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the trigger delay in seconds.
Default is 0 s.
Description
Selects the Run Stopped softkey as a trigger source.
Selects any configured digital IO (with Trigger Input
function) as a trigger source
Selects a remote interface command as a trigger
source.
103