From the remote interface:
To configure the load operating mode to CV mode:
FUNC VOLT, (@1)
To set the sequencer to Pulse mode:
TRAN:MODE PULS, (@1)
To configure main voltage level with 5 V:
VOLT 5, (@1)
To configure the value for the transient voltage level with 10 V:
VOLT:TLEV 10, (@1)
To configure the transient pulse width with 0.5 s :
TRAN:TWID 0.5, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
VOLT:MODE LIST
To enable the input channel:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)
Toggle mode
Applicable for Load mode only.
Toggle mode generates a repetitive pulse stream that toggles between two load levels. It causes the load input to
alternate between two pre-defined levels as in continuous operation except that the transient points are controlled
by explicit triggers instead of an internal transient generator.
Step 1– Configure the sequence properties
Configure the Sequencer Toggle accordingly. Refer to the below table for details.