Note that you will only need to configure the Turn-on delays, not the turn-off delays.
The turn-on delays for the input channels are as follows:
– Input 1: 0.1 s
Step 4 - Configure the Scope View traces
– Uncheck V1.
- Check I1.
– Press Knob Scale and use the Vertical knob to set I1 to 1 A/Div.
– Press Knob Ofs and use the Vertical knob to set offset to 1 A.
– Press Knob Scale and use the Horizontal knob to set the timebase to 1 s.
Step 5 – Configure the Scope properties:
Press Properties to configure the scope properties as follows:
– Press Settings > Trigger Source > List Key to select [List Run/Stop] key as a trigger source. Press Back to return
to Settings menu.
– Press Trigger Mode > Auto for auto sweep measurement. Press Back twice to return to the Scope Properties
menu.
– Set the Time Reference to Left in Horizontal Settings (Hor) window: Press Settings > Time Ref Lf.