Under Voltage Inhibit
Applicable for Load mode only.
When the under-voltage inhibit function is enabled by selecting a mode, the load will not sink current until the input
voltage rises above the voltage-on setting.
Press UVI Mode to specify the under-voltage inhibit mode:
– Off: Turns the under-voltage inhibit function off.
– Live: Turns the input off whenever the voltage drops below the voltage-on setting. Turns the input back on when
the voltage reaches the voltage-on setting.
– Latched: Lets the load sink current when the voltage subsequently drops below the voltage-on setting. An
under-voltage inhibit condition is annunciated by the UVI status bit.
Under-voltage inhibit is not available when loads are grouped or when the unit is operating in CV mode.
To set the voltage-on setting, enter the voltage-on value in the Voltage On field. The voltage-on value can be
programmed from 0.02 to 61.2 V.
Clears OVP, OCP or OPP Event
To clear the protection function, first remove that condition that caused the protection fault.
Press Protection Clear to clear the protection function and returns the output to its previous operating state.
From the remote interface:
To set the over-voltage protection to the maximum limit:
VOLT:PROT MAX
To enable the over-current protection:
CURR:PROT:STAT ON, (@1)
To set the over-current protection delay time at 0.1 seconds:
CURR:PROT:DEL 0.1, (@1)
To set the over-current protection delay timer start to CC transition:
CURR:PROT:DEL:STAR CCTR, (@1)
To enable the over-power protection:
POW:PROT:STAT ON, (@1)
To set the over-power protection delay time at 0.2 seconds:
POW:PROT:DEL 0.2, (@1)
To set the over-current protection delay timer start to CC transition:
CURR:PROT:DEL:STAR CCTR
80