1. Configure the Rise Rate or Fall Rate accordingly. Use the navigation keys to select the field; use the numeric
entry keys to enter the value. The value is set when you press [Enter]. You can also enable the Max checkbox to
set the slew rate to the maximum value.
2. Select VOLT Rise On or VOLT Rise Off to enable or disable the voltage slew rate rising configuration and select
VOLT Fall On or VOLT Fall Off to enable or disable the voltage slew rate falling configuration.
3. Press Back to exit and return to the meter-view display.
From the remote interface:
To set output to 5 V and 8 A:
APPL CH1 5, 8
To set the remote sense relay to 4-wire sense:
VOLT:SENS EXT, (@1)
To set the preferred mode to Voltage:
OUTP:PMOD VOLT, (@1)
To set the preferred mode to Current:
OUTP:PMOD CURR, (@1)
To set the rising voltage slew rate to 5 volts per second:
VOLT:SLEW:RIS 5, (@1)
To set the falling voltage slew rate to the maximum value:
VOLT:SLEW:FALL MAX, (@1)