Settings
Available Key Set-
tings
Current Mode
Fix, Stp, or Lst
Voltage Mode
Resistance Mode
Power Mode
Trigger Source
Key, IO, or Rmt
Trigger Delay
Load mode:
0 to 0.255 s
Power Supply mode:
0 to 3600 s
Repeat Count
1 to 9999
Continuous
-
Step 3 – Run the output sequence list
When Sequencer List is initiated, all the List properties cannot be configured, which includes adding or remov-
ing step from the list.
When Step mode transient system is initiated, properties such as voltage mode, current mode, resistance
mode, power mode, trigger source, and trigger delay cannot be configured.
If both Voltage and Current Mode is set to Fix and Trigger Source is set to Key, pressing Run will automatically
set both voltage and current mode to List and start the list operation.
96
Sets the current/voltage/resistance/power mode. This determines what happens to the
output current/voltage/resistance/power when the system is initiated or triggered.
Default is Fix.
Action required:
Press Curr Mode, Volt Mode, Res Mode, or Pow Mode to toggle between Fix, Stp, and Lst.
Fix (Fixed)
Keeps the output at its immediate value.
Stp (Step)
Steps the output to the triggered level when a trigger occurs
Lst (List)
Causes the output to follow the list values when a trigger occurs.
Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the trigger delay in seconds.
Default is 0 s.
Sets the list repeat count. This sets the number of times that a list is executed before it
completes.
Default is 1.
Enables check box to repeat the list continuously.
Description
Selects the Run Stopped softkey as a trigger
source.
Selects any configured digital IO (with Trigger
Input function) as a trigger source
Selects a remote interface command as a trig-
ger source.
Keysight E36731A User's Guide