Settings Sample Points
Time Reference
Trigger Source
128
256k, 128k, 64k, 32k,
Specifies the number of points in a scope trace. Default is 4096.
16k, 8192, 4098,
Action required:
2048, or 1024
Press Sample Points to view and select the nine available sample points.
Lf, Ctr or Rg
Specifies the reference point on the display.
This is the position of the trigger if no offset has been set.
Default is Lf.
Action required:
Press Time Ref to toggle between Lf, Ctr and Rg.
Lf (Left)
Ctr (Center)
Rg (Right)
V1 Level, I1 Level, List
Configures the trigger source.
Key, On/Off, DIO or
Default is Key.
Remote
Action required:
Press Trigger Source to select a trigger source. This trigger source will trigger the
scope measurements on all input channels.
V1 Level
I1 Level
List Key (List Run-
/Stop Key)
On/Off (Input
On/Off Key)
IO (DIO Trigger In) Selects any configured digital IO (with Trigger Input function)
Remote (Remote
Command)
Lets you see the waveform after the trigger event.
Lets you see the waveform before and after the trigger event.
Lets you see the waveform up to the trigger event
Selects the voltage trigger level as trigger source.
Triggers the measurement when the voltage of the
corresponding input passes through the specified level.
Selects the current trigger level as trigger source.
Triggers the measurement when the current of the
corresponding input passes through the specified level.
Selects the List Run/Stop key as a trigger source.
Selects the Input On/Off keys as a trigger source.
as a trigger source.
Selects a remote interface command as a trigger source.
Keysight E36731A User's Guide