To configure the transient duty cycle with 10.5%:
TRAN:DCYC 10.5, (@1)
To configure trigger delay with 0.2 s:
TRIG:TRAN:DEL 0.2, (@1)
To configure the transient repeat count with 20:
TRAN:COUN 20, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
VOLT:MODE LIST
To enable the input channel:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)
Pulse mode
Applicable for Load mode only.
Pulse mode generates a load change that returns to its original state after some time period.
It is similar to continuous operation with the following exceptions:
– To get a pulse, an explicit trigger is required.
– One pulse results from each trigger. Therefore, frequency cannot be programmed. The pulse width is pro-
grammable.
Step 1– Configure the sequence properties
Configure the Sequencer Pulse accordingly. Refer to the below table for details.
100
Keysight E36731A User's Guide