Scope View
Press [Scope/Datalog] key to view the scope. This key toggles between the Scope view and Data Logger view.
Items
1
Trace controls
2
Data bar
3
Time/Div
4
Offset time
5
Sequencer status
6
Trigger mode
7
Trigger Source
Amplitude
8
Sample period
9
Range
10
Scope status
11
Trigger level
122
Identifies the voltage or current that will be displayed. √ indicates the trace is on. Dashes (----) indic-
ate the trace is off. Select the trace and press [Enter] to turn it on or off.
The data bar represents all of the logged data. The yellow part indicates the portion of the data that is
visible on the display. The black part represents the data that is not visible.
Identifies the horizontal time-base setting. This can be adjusted using the Horizontal knob after press-
ing Knob Sc.
Indicates the time from the trigger point indicator to the vertical center line of the grid. Negative
values indicate the center line is to the left of the trigger point. Positive values indicate the center line
is to the right of the trigger point. The trigger point can be adjusted using the Horizontal knob after
pressing Knob Ofs.
Indicates the Sequencer is running, or waiting for a trigger. No indicator when Sequencer is idle.
Indicates the trigger mode (Auto, Single, or Triggered).
In the figure, the trigger source is a voltage level on input 1.
Up trigger indicates the measurement will be triggered on the up-slope (positive).
Down trigger indicates the measurement will be triggered on the down-slope (negative).
If the trigger source is set to a voltage or current level, the amplitude of the trigger level is indicated
beside the trigger source. In the figure, the voltage trigger level is set to 3.152 V.
The indicated scope sample period is based on the horizontal time/div. setting. When the time/div.
setting is less than 2 ms/division, the scope will sample at its fastest rate, 5.12 µs.
Range indicates the measurement range setting of the selected trace.
Indicates whether the scope is Running, Stopped, or Waiting for a trigger.
Shows the location of the voltage or current trigger level and input. In this example, the voltage trig-
ger level of input 1 is shown. The trigger source and amplitude are shown at the left corner of the dis-
play.
Description
Keysight E36731A User's Guide