Table 3
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity
IEC 60601 Test Level
Test
Conducted
3 V
RF
0.15 MHz to 80 MHz
IEC 61000-4-
6 V in ISM bands between 0.15
6
MHz and 80 MHz
Radiated RF
10V/m
IEC 61000-4-
80 MHz to 2.7 GHz
3
Table 4
Guidance and Manufacturer's Declaration - IMMUNITY to proximity fields from RF
wireless communications equipment
Immunity
IEC60601 Test Level
Test
Test
Frequency
Radiated
385 MHz
RF
IEC 61000-
4-3
450 MHz
710 MHz
745 MHz
780 MHz
810 MHz
870 MHz
930 MHz
1720 MHz
1845 MHz
1970 MHz
Modulation
Maximum
Power
**Pulse
1.8W
Modulation:
18Hz
*FM+ 5Hz
2 W
deviation:
1kHz sine
**Pulse
0.2 W
Modulation:
217Hz
**Pulse
2 W
Modulation:
18Hz
**Pulse
2 W
Modulation:
217Hz
E-4
Compliance Level
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15
MHz and 80 MHz
10V/m
Immunity
Level
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
Compliance
Level
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m