Warning
To avoid the risk of electric shock, when the device is used together with the
recorder base, be sure that the recorder base must only be connected to a
supply mains with protective earth.
12.1.2 Defibrillator and Pacemaker Warnings
Warning
When used with a defibrillator or pacemaker, all electrodes connected and not
connected to the patient and the patient should not be grounded.
Warning
Before defibrillating, make sure the patient is completely isolated and avoid
touching any metal part of the device in case of electric shock.
Warning
Before defibrillating, remove all electrodes, gel or cloth from the patient in case
of any possible burnt. When the electrode paddle of defibrillator is in direct
contact with these materials, the discharge capacity will cause severe electric
burn of patients.
Warning
Before defibrillating, enable the ADS function and select 0.67Hz filter.
Warning
Use patient cable with defibrillator protection specified by the manufacturer
while defibrillating. Otherwise there might be electric burnt of the patient or
damage of the device. After defibrillation, under the standard sensitivity setting,
the ECG waveform will return to 80% of the normal amplitude within 5 seconds.
Warning
During defibrillation, use disposable electrodes and ECG adapter wires specified
by the manufacturer and use them in accordance with their instructions for use.
Warning
After defibrillation, the ADS filter is set at 0.67Hz, and the cardiogram is
displayed and maintained within 10 seconds.
Warning
Use only the patient cable and electrodes supplied by the manufacturer while
defibrillating.
11-4