14.3.9 Inspections Before Power-On
To ensure the safe and effective operation of the device, perform the
following inspections before power-on and operation.
Operating Environment:
Make sure that there is no electromagnetic interference source around
the equipment, such as electrosurgical device, ultrasonic diagnostic
device, radioactive device, etc. Switch off these devices when necessary.
Power Supply:
Check that the batteries are installed in the device and the recorder base.
Make sure that the batteries are fully charged.
Check that the power adapter is securely connected if the device is
powered by DC power supply.
Check that the power cord is securely connected if the recorder is
powered by AC power supply. Use only power socket that is properly
grounded.
Patient Cable:
Make sure that the patient cable is connected to the device firmly.
Electrodes:
Make sure that all electrodes are connected to lead wires of the patient
cable correctly. Ensure that the electrodes, especially the chest electrodes
do not contact with each other.
Patient:
The patient's hands and feet should not come into contact with
conducting objects such as the metal part of the bed.
Ensure that the patient is warm and relaxed, and breathes calmly.
14.3.10 Login and Account Management
Power-On and Login
Press the Power On/Off key to turn on the device, it will enter the startup
screen, and then enter the login screen. The login method depends on the
diagnostic mode configured by the device.
Auto diagnosis mode
You need to input the correct password (initial password: 135790) to enter
the main screen of the device.
AI diagnosis mode
11-6