When the HR exceeds the detectable HR range, the HR value area is
displayed as "-".
0 means cardiac arrest, displayed as 0.
When all leads / rhythm leads fall off, the HR will be displayed as "-"
by default.
4
Lead indication area
Display the lead signal quality detection indicator:
Green: lead connection is normal, waveform quality is good and free
from interferences.
Orange: lead connection is normal, but waveform is interfered.
Yellow: the lead is off.
Click the [
connection status in the popup window. The name and position of
the electrode that fell off are displayed in yellow, and that of the one
not fell off are displayed in green.
5
Prompt information area
Display prompt information such as "All leads off", "HR overrange".
6
Status display area
Display the current network, internal battery, external power and external
USB device connection status of the device.
Wired network
: indicates that the device is connected to a wired network
successfully. Not displayed when not connected.
Wireless networks
: indicates that a WiFi wireless network is connected. The solid
part indicates network signal strength. Not displayed when not
connected.
: indicates that the mobile network is connected and displays the
name of the operator to which it belongs. The solid part indicates
network signal strength. Not displayed when not connected.
Battery
If a battery is installed, the percentage of remaining battery power and
the battery icon will be displayed; otherwise, it will not be displayed.
] icon to view the electrode connection diagram and
: indicates that the battery is being charged.