Annexe E EMC and Radio Regulatory Compliance
E.1 EMC Compliance
Basic performance: The device can acquire ECG data normally.
Warning
Don't use near active HF surgical equipment and the RF shielded room of an ME
system for magnetic resonance imaging, where the intensity of EM disturbances
is high.
Warning
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally.
Warning
Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Warning
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the device, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this device could result.
Warning
The device may still be interfered even if other devices meet the emission
requirements of the corresponding national standards.
Caution
Users shall install and use the device according to the EMC information provided
in this manual.
Caution
Mobile or portable RF communication equipment may affect the performance of
the device. Avoid strong electromagnetic interference when in use, such as near
mobile phones, microwave ovens, etc.
E-1