Frequency-domain analysis index: Total Power (ms*ms), VLF
(extremely low frequency, ms*ms), LF (low frequency, ms*ms),
LFnorm (nu), HF (high frequency ms*ms), HFnorm (nu), LF/HF
RR Histogram
RR Interval Diff Histogram
Frequency Chart
17.4.2 AI Diagnosis
In the AI Server login mode, after the ECG waveform is stable, click the
[Start] button, the device starts recording the ECG waveform. If you have
enabled [Auto Upload] from the [ECG Setup] screen, ECG data is
automatically uploaded to the connected AI server for analysis at the
completion of ECG acquisition. After the server returns the diagnosis
report, the status of the relevant patient record in the file management
menu will change to "Diagnosed", you can view the diagnosis result and
print the report.
If the auto upload function is not enabled or the upload fails, ECG data will
be automatically stored in the device, and they can be manually uploaded
to the server in the file management menu.
17.5 Printing Reports
You can print ECG reports through an external printer. Refer to 3.3.11
Connecting the Printer for methods for connecting the printer to your
device. Before printing a report, check that the paper is properly loaded.
To load the paper for the external printer, refer to the printer's
accompanying instructions for use.
The 10.1" device can also be configured with a recorder base to output
the ECG reports. Before printing a report, refer to 3.3.3 Loading the
Recording Paper for loading paper for the recorder base.
Note
If the [Preview] option in the [ECG Setup] screen is disabled, the device
automatically prints the ECG report after the ECG data is acquired and analyzed.
Note
If the [Printout] option in the [Record Setup] screen is disabled, click the [Start]
button to save but cannot print the ECG report.
You can also print reports by exporting them to a USB flash disk.