21.2.2 Cleaning Patient Cable and Electrodes
Before cleaning the patient cable and electrodes, remove the patient
cable from the device.
For the cleaning of the patient cable and electrodes, refer to their
instructions for use delivered with the accessories.
21.2.3 Cleaning the Thermal Print Head
Dirty print head deteriorates the printing quality. Clean the print head at
least once a month or as needed.
To clean the thermal print head, follow the steps below:
1.
Turn off the recorder base.
2.
Open the paper compartment cover and take out the recording
paper.
3.
Wipe the print head gently with a clean soft cloth dampened in 75%
alcohol. For stubborn stain, soak it with a small amount of alcohol
first and wipe it off with a clean soft cloth.
4.
Reload the recording paper and close the paper compartment cover
after the print head is completely air dried.
21.3 Disinfection
Disinfection of main unit of the device is not necessary. To avoid
permanent damage to the device, disinfection can be performed only
when it has been considered as necessary according to your hospital's
regulations. Before disinfection, clean the device first.
For the disinfection of the patient cable and electrodes, refer to their
instructions for use delivered with the accessories.
21.4 Care and Maintenance
To ensure the performance and safety of the device and its accessories,
routine care and maintenance should be carried out.
21.4.1 Device and Recorder Base
Follow the below guidelines to maintain the device and recorder base:
Avoid excessive temperature, sunshine, humidity and dirt. Prevent
shaking it violently when moving it to another place.
Prevent any liquid from penetrating into the device, otherwise the
safety and performance of the device cannot be guaranteed.