Télécharger Imprimer la page

Spengler LeECG Serie Manuel De L'utilisateur page 217

Publicité

Les langues disponibles

Les langues disponibles

Table 2
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity Test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 Test Level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s) to
earth
0 % UT; 0.5 cycle At 0°, 45°,
90°, 135°, 180°, 225°, 270°and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
30 A/m
E-3
Compliance Level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV
air
±2 kV for power supply
lines
±1 kV for input/output
lines
± 0.5kV, ± 1 kV line(s) to
lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
0 % UT; 0.5 cycle At 0°, 45°,
90°, 135°, 180°, 225°,
270°and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
30 A/m

Publicité

loading

Ce manuel est également adapté pour:

Neoecg serieNeoecg s120Neoecg t120Leecg os12Leecg ot12