No.
Messages or Troubles
13
Recorder out of paper
14
Recorder door not closed
15
XX lead off (All leads off)
16
Low battery!
Battery depleted. Power off
17
soon.
18
Export failed
19
Insufficient memory space
Some Lead without Waveform
20
Printout
Prompt Messages and Troubleshooting
Solutions
Load the paper properly.
Close the recorder door.
1. Check the corresponding
electrodes and lead wires. Re-apply
the electrodes or reconnect the lead
wires if necessary.
2. Check that the patient cable is
properly connected to the device.
3. If you have done all of the above
steps, but still no waveforms, check
that whether the device is
inoperable due to an overload or
saturation of any part of the
amplifier.
Charge the battery immediately.
Connect the DC power supply or use
the recorder base to power the
device, or charge the battery
immediately.
Check the USB device and export
data again.
Delete unwanted historical files or
change the storage device / location.
If you acquire the ECG data
immediately after the leadwires are
applied to the patient, the ECG
traces may not display because the
ADS is not stable yet. Normally it is
necessary to wait for the waveform
of each lead to be stable if all leads
are in good contact before ECG
measurement.