Live image vs. recorded image
Round BLUE buttons are shown in the blue Live Image tab
Square YELLOW or GREEN buttons are shown in the yellow File Management tab
green Settings tab
Battery status: During start up, aView powers up and configures the visualization device.
If the aView battery icon on the screen changes from fully charged to low battery (red battery)
within 30 minutes, aView must be replaced.
Connection to external monitor
The image from an Ambu visualization device can be viewed on an external monitor using the
video out interface on aView. Connect the external monitor to the composite interface on the
right side of aView using the adaptor cable supplied (refer to the online aView User Guide on
www.ambu.com). Please consult the manual of the external monitor for further information on
how to connect an external video source by composite.
4.4. After Use
Final steps
1.
Disconnect the Ambu visualization device from aView. For disposal of the visualisation
device, refer to the Instruction for Use of the specific device.
2.
Switch off aView by pressing the power button
hourglass indicating that aView is powering down will appear on the screen and
aView will power down.
3.
Clean and disinfect aView (cf. section 5).
4.
If the battery level of aView is low, charge the aView (cf. section 4.1).
5. Cleaning and Disinfection of aView
The aView should be cleaned and disinfected before and after each use. It is recommended that
aView is cleaned and disinfected before and after use following the instructions below. Ambu
has validated these instructions in accordance with AAMI TIR 12 & 30. Any deviation from the
instructions should be properly evaluated for effectiveness and potential adverse consequences
by the reprocessor, to ensure that the device continues to fulfil its intended purpose.
Clean and disinfect the aView according to good medical practice using one of the procedures below:
Procedure 1 - SANI-CLOTH® BLEACH from PDI
Cleaning
Use a wipe to remove heavy soil. All blood and other body fluids must be thoroughly
cleaned from surfaces and objects before disinfection by germicidal wipe.
Disinfecting
1.
For heavily soiled surfaces, use a wipe to pre-clean aView prior to disinfecting.
2.
Unfold a clean wipe and thoroughly wet the surface of aView.
3.
The treated surfaces must remain visibly wet for a full four (4) minutes. Use additional
wipes if needed to assure continuous 4 minutes wet contact time.
4.
Let aView air dry.
Procedure 2 - SUPER SANI-CLOTH® from PDI
Cleaning
Use a wipe to remove heavy soil. All blood and other body fluids must be thoroughly
cleaned from surfaces and objects before disinfection by germicidal wipe.
Disinfecting
1.
For heavily soiled surfaces, use a wipe to pre-clean aView prior to disinfecting.
2.
Unfold a clean wipe and thoroughly wet the surface of aView.
3.
The treated surfaces must remain visibly wet for a full two (2) minutes. Use additional
wipes if needed to assure continuous 2 minutes wet contact time.
4.
Let aView air dry.
10
and indicates a recorded image.
and indicates a live image.
for at least 2 seconds. A blue
and the