EN
29
Key switch
Only authorized skilled personnel may operate the device in the "TEACH" key setting.
The key is to be kept in a safe place so that it is protected from unauthorized access.
The following program locks and functions can be set with the key switch when the rotor is at a standstill:
Key setting
Program locking / function
The centrifuge can be operated via the control panel.
0
The centrifuge cannot be controlled via the interface.
LOCK 2 is displayed in the "
LOCK 2
The centrifuge can be controlled via the interface.
The centrifuge cannot be operated via the control panel.
TEACH
Display e.g.
Teach
PROG
= : Position of the hatch
RPM
PROG
30
Setting rotor position 1
Only authorized skilled personnel may set rotor position 1.
Every time the rotor is installed, rotor position 1 must be reset afterwards. Rotor position 1 is the position where place
1 of the rotor must be located for loading and unloading. Rotor position 1 can be set as follows when the rotor is at a
standstill:
•
Look through the window in the lid to make sure that the rotor is at a standstill.
•
Turn the key to the "TEACH" setting. For example, Teach Open =Close might be displayed.
•
Keep the
key pressed until the hatch is completely open. Teach =Open Close is displayed.
•
Press the
key. The saved rotor position 1 is displayed, e.g. Teach Pos 1 : 1812 <Start=ok>.
PROG
•
Turn the rotor slowly until you hear an acoustic signal. Afterwards, continue to turn it until place 1 of the rotor is in
the desired loading/unloading position. Hold the rotor in this position.
•
Press the
button to save the setting.
START
As a confirmation, Store Pos 1 ... is briefly displayed, and then e.g. Goto Pos 1/24 2826
If "no zero pulse !" is briefly displayed, the rotor was turned to the loading/unloading position before the
acoustic signal sounded. In this case, repeat the rotor position 1 setting.
•
To test the positioning, set the desired rotor place with the rotary knob , and then press the
rotor continues to turn until the set rotor place is in the loading/unloading position.
•
Press the
key to exit the menu. Teach =Open Close is displayed.
STOP / OPEN
•
Keep the
key pressed until the hatch is completely closed. Teach Open =Close is displayed.
RPM
•
Turn the key back to the "LOCK 2" position.
48/108
" field.
Open
=Close
RPM
: Open the hatch (keep key pressed)
: Close the hatch (keep key pressed)
: Set rotor position 1. See the chapter "Setting rotor position 1".
0.
key. The
START