EN
20
Displayed centrifugation data after switch-on.
After switch-on the centrifugation data from program 1, or from the last program that was used, is displayed.
If the rotor is at standstill, this can be set as follows:
•
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
•
Press the
key as often as necessary until the ->Settings is displayed.
PROG
•
Press the
key. The SOUND / BELL = off/on is displayed.
START
•
Press the
key as often as necessary until Start program = Last/First is displayed.
PROG
•
Use the adjusting knob
Last = the program used last, First = program 1.
•
Press the
key to save the setting.
START
Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.
•
Press the
key once to leave the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
21
Set temperature unit
The temperature can be entered in degrees Celsius (°C) or in degrees Fahrenheit (°F).
When the rotor is at standstill, the temperature unit must be set as follows:
•
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
•
Press the
key as often as necessary until the ->Settings is displayed.
PROG
•
Press the
key. The SOUND / BELL = off/on is displayed.
START
•
Press the
key as often as necessary until Temp Unit = Celsius/Fahrenheit is displayed.
PROG
•
Use the adjusting knob
•
Press the
key to save the setting.
START
Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.
•
Press the
key once to leave the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
22
Centrifuge address
Ex works, the address is ] = 29. Address is set.
23
RCF integral and B ramp functions
These functions are not activated as standard.
24
Query of operating hours and the number of centrifugation runs
The operating hours are divided up into internal and external operating hours.
Internal operating hours: Total time the device was switched on.
External operating hours: Total time of previous centrifugation runs.
The operating hours and the number of centrifugation runs can be queried as follows when the rotor is at standstill:
•
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
•
Press the
key as often as necessary until the -> Operating Time is displayed.
PROG
•
Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START
•
Press the
key. The internal operating hours are displayed, e.g. OP Time int = 1h36m.
PROG
•
Press the
key. The number of centrifugation runs is displayed, e.g. Number of Starts = 10.
PROG
•
Press the
key to leave the menu, "Operating Time" or press the key
OPEN / STOP
leave the "Machine Menu".