13.3

Time area

This area contains the running time display and the
preselected or changed during operation.
When stationary, the preselected switch-on time is displayed, as well as the remaining run time during operation. The
decimal point flashes each second during operation.
At the end of operation, the display flashes each second until the lid is opened or a key is activated.
13.4

Key area

Start key
To start a run with the preselected parameters.
START
.
Stop key
The STOP key allows a centrifuging run to be stopped at any time.
STOP
.
Pulse key
Brief centrifuging whilst the key is pressed down.
The run time is displayed in seconds in the run time display.
After 60 seconds, the run time is displayed in minutes.
Lid key
After standstill the lid can be unlocked by pressing the corresponding key.
14

Adjustable parameters

Speed

Adjustable up to max. speed of the rotor used.
Time
or
"—"
or
Braking force
15

Preselecting centrifuging parameters or changing them during operation

15.1
Speed
Each time the
and
arrow keys are activated, the speed is changed by 100 rpm. If an arrow key is kept pressed
down for longer, the speed is changed at an increasing rate.
If the speed is changed during operation, this becomes effective immediately, i.e. the speed is adjusted.
and
min. speed 500 rpm
max. speed 14000 rpm
in steps of 100 rpm
min. preset time 1 min
max. preset time 99 min
in steps of 1 min
continuous operation
pulse ( short-time )
full braking effect
low braking effect
arrows, with which the switch-on time can be