1
Use according to specification
The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG.
The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also
includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this
purpose. A different use or application over and above this is deemed not in accordance with the specifications. The
company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
If the centrifuge will be built into another machine or will be integrated in a system the manufacturer of the complete
system is responsible for its security.
2
Residual risks
The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3
Technical specifications
Manufacturer
Model
Type
Mains voltage (± 10%)
Mains frequency
Connected load
Cooling medium
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
− Set-up site
− Altitude
− Ambient temperature
− Humidity
− Excess-voltage category
(IEC 60364-4-443)
− Pollution degree
Device protection class
EMC
− Emitted interference,
Interference immunity
Noise level (dependent on rotor)
Dimensions
− Width
− Depth
− Height
Weight
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
ROTANTA 460 Robotic
5680-70,
5680-78,
5680-70 (H)
5680-78 (H)
230-240 V 1∼ 200-220 V 1∼
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN / IEC 61326-1, Class B;
/
* mm (* with cables connected)
697
762
5680-75,
5680-71,
5680-75 (H)
5680-71 (H)
127 V 1∼
110-120 V 1∼
50-60 Hz
max. 1800 VA
R 404A
4 x 750 ml
3
1.2 kg/dm
6200
5845
40000 Nm
yes
Indoors only
5°C to 35°C
ΙΙ
2
Ι
FCC Class A
≤ 62 dB(A)
mm
554
723 mm
approx. 159 kg
EN
5680-74,
5680-74 (H)
100 V 1∼
33/108