3 INSTALLATION
3.2.6 Electrical wiring, compact version, 115/230 VAC, without relay
Before wiring the device, please read § 3.2.1 and 3.2.3 carefully.
Lift the transparent lid after having unfastened the screw. Remove the cover of the device by
unscrewing the 4 screws, pass the cables through the cable glands and connect according
to the pin assignment below.
Always seal the unused cable gland using the supplied obstructor to
ensure the tightness of the device.
Unscrew the cable gland nut, insert the obstructor and screw the nut back
onto the cable gland.
Only manipulate the switch when the device is not powered.
Red wire * L+
Strap to be removed if
4-20 mA output is used
* wired ex-works
Fig. 3.11
Wiring compact transmitter, 115/230 VAC, without relay
Black wire * L-
Yellow wire* PE
Not used
Conductivity sensor
cable
E-24-
CONDUCTIVITY 8225
115 VAC or 230 VAC
power supply selection
T 125 mA
L
N
PE
Fuse
115/230 VAC power supply
connection