3 INSTALLATION
CONDUCTIVITY 8225
3.1.2
Panel-mounted version
Respect the installation specifications of the conductivity sensor described in § 3.1.1 or in
the corresponding instruction manual then follow the instructions below to install the device:
1. Use the supplied drilling jig to cut away the necessary opening in the panel. Make sure to
scrupulously respect the dimensions indicated.
2. Place the gasket on the 4 screws of the cover.
Note: Use the 4 M4x25 screws supplied if the panel door is too thick.
3. Place the cover+gasket set on the cut-away with the electronic board facing the inside of
the panel.
4. Insert the 4 washers onto the 4 screws and fit the device to the panel using the 4 nuts.
5. Connect according to instructions described in chapter 3.2.
6. Use the supplied 4 cable clips to fit the cables to the protective plate (see § 3.2.3).
Screw
Washer
Nut
Cable clip
Gasket
Fig. 3.2
Mounting 8225, panel-mounted version
E-14-