4 OPERATION
After, the device will be immersed into the solution (temperature < T- or < 25 °C if T- > 25
°C) and the solution is heated up. The temperature compensation coefficients will automati-
cally be determined until T+ or 25 °C (if T+ < 25 °C) is reached. The values are memorized
and can be called up any time by selecting "SPECIAL" in the "AUTO" option.
T ° COEFF
ENTER
CURRENT
CURRENT
WARNING: The temperature increase must
be slow in order to compensate the thermal
resistance of the sensor. Also avoid bubbles
appearing on electrode surface. The user
can quit the Teach-In function at any time by
pressing
for 5 sec.
0......9
If the message "ERROR" is displayed, a pro-
blem has occured during Teach-In option
(for example, the liquid gas been heated
too quickly). In this case, the operation must
be repeated.
CURRENT
LINEAR
ENTER
ENTER
ENTER
AUTO
ENTER
TEACH IN
ENTER
ENTER
E-37-
CONDUCTIVITY 8225
TC= 00.00
0......9
TC= 02.10
NAOH
The coefficients in
the "SPECIAL" opti-
HNO3
on equal zero until
a Teach-in function
H2SO4
has been comple-
ted, i.e. there is
NACL
no temperature
compensation of
SPECIAL
the conductivity.
H2O PURE
T-= 000
0......9
T-= 015
ENTER
MEASURE
ENTER
The message "MEASURE"
T °= 065
appears only when tem-
perature has reached T-.
MEAS END
T+= 000
0......9
T+=110