Fitting The Third Column Element - Seca 286 Mode D'emploi

Masquer les pouces Voir aussi pour 286:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 246

Fitting the third column element

3. Tighten up the two hex socket bolts to fix the multifunctional display in the
desired position.
4. Route the power supply connection cable of the multifunctional display in
the groove of the column elements as far as the lower part of the device.
5. Draw the power supply connection cable through the bore of the lower
part of the device.
ATTENTION!
Malfunction as a result of assembly error
If the cables are fitted under severe mechanical load, faulty displays
and failure of the display may result.
Lay all cables so that they are not excessively bent and so that
connectors are not snapped off.
6. Wind the free end of the power supply connection cable onto the cable
store in the lower part of the housing.
7. Plug the connector of the power supply connection cable into the
appropriate socket of the weighing platform.
8. Fix the power supply connection cable in the groove of the column
elements using the two cable clips.
9. Swivel the multifunctional display so that is convenient for you to read.
ATTENTION!
Device damage as a result of assembly error
If the long and short threaded rods are switched, it will not be possible
to assemble the device so that it is fit for purpose.
Ensure that the long threaded rod is fitted when assembling the third
column element (see "Scope of delivery" on page 210).
To fit the third column element, proceed as outlined below.
Before you really get started ... • 213