Problem
... the display "Er :6 :82" appears?
... the display "Er :6 :83" appears?
... the display "Er :6 :84" appears?
10.MAINTENANCE
11.TECHNICAL DATA
11.1 General technical data
Dimensions
• Depth
• Width
• Height
Net weight
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
Ambient conditions, transport
• Temperature
• Air pressure
• Humidity
Height of digits
• Multifunctional display, three-line
An error occurred during the measuring operation.
- Ask the patient to keep still and repeat the measuring operation
• If the error message recurs:
- inform seca Service
An error occurred during the reference measurement.
- Step off the platform
- Check that there are no objects or people in the immediate vicinity of the
device (see "Setting up the device" on page 215)
• The ambient temperature of the device is too high or too low.
- Set up the device in an ambient temperature between +10 °C and
+40 °C
- Wait for around 15 minutes until the device has adapted to the ambient
temperature
• The temperature sensor is faulty.
- Inform seca Service
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor servicing
►
Have servicing and repairs carried out exclusively by seca service or
by an authorized service partner.
►
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
General technical data
Cause/solution
+10° C to +40° C (50 °F to 104 °F)
30 % to 80 %, no condensation
-10° C to +65° C (14 °F to 149 °F)
0 % to 95 %, no condensation
-10° C to +65° C (14 °F to 149 °F)
0 % to 95 %, no condensation
English
466 mm
434 mm
2270 mm
16.5 kg
700 hPa - 1060 hPa
700 hPa - 1060 hPa
700 hPa - 1060 hPa
14 mm
Maintenance • 243