Install the third column
element
Installing the push-in
To connect the column elements, proceed as follows:
1. Screw a threaded rod hand-tight into the elongated
nut of the second column element.
2. Push the column element over the threaded rod
onto the upright connector of the second column
ELEMENT.
3. Position the end cap on the second column
element so that the threaded rod protrudes from
the hole of the end cap.
4. Place the plain washer on the free end of the
threaded rod.
5. Place the nut on the free end of the threaded rod.
6. Tighten the nut.
7. Place the decorative cap on the nut.
ATTENTION!
Incorrect measurements due to installation
scale
errors
If the push-in scale is installed incorrectly the
head slide cannot determine any usable
readings.
− Align the push-in scale so that the printing
− Align the push-in scale so that the area which
− Ensure that the push-in scale latches into
♦ Push the push-on scale into the groove in the side
of the column elements until the scale latches into
position under the end cap.
remains visible when the scale is pushed into
the groove.
shows the product number forms the bottom
end.
position under the end cap.