• Do not use i i B B e e d d Awareness to replace your patient monitoring protocol.
• Do not use i i B B e e d d Awareness as a lock indicator for siderails. i i B B e e d d Awareness only detects the position of the siderails.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always make sure that all persons and equipment are away from the area below and around the Fowler before you
activate the CPR release. The CPR release is for emergency use only.
• Always follow hospital protocol to determine the use of restraint straps and restraint strap locations. Stryker is not
responsible for the type or use of restraint straps on any Stryker product.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
• Always raise the siderails when the litter is in its full down position. This prevents the scale system from weighing a
patient inaccurately.
• Always make sure that the siderails are locked before you arm i i B B e e d d Awareness.
• Always make sure that you set the desired product parameters before you arm i i B B e e d d Awareness.
• Do not use accessories that cover the footboard LED light bars.
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the patient from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
• Always clean V V e e l l c c r r o o
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
• Always clean V V e e l l c c r r o o ® after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
• Do not use quaternary disinfectants formulated with glycol ethers.
EN
4
® ®
after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
3005-009-005 Rev AA.1