W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n
The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
W W A A R R N N I I N N G G
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
C C A A U U T T I I O O N N
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
N N o o t t e e - - Provides special information to make maintenance easier or important instructions clearer.
S S u u m m m m a a r r y y o o f f s s a a f f e e t t y y p p r r e e c c a a u u t t i i o o n n s s
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
W W A A R R N N I I N N G G
• Always plug the product directly into a grounded, hospital grade wall outlet. You can only achieve grounding reliability
when you use a hospital grade wall outlet. This product is equipped with a hospital-grade plug for protection against
electric shock hazard.
• Always use a Stryker supplied interface cable. Use of any other cable may cause the bed to not function as intended,
which may result in patient or user injury.
• Always associate or map the i i B B e e d d L L o o c c a a t t o o r r to the room or location to provide accurate location information. If you move
an i i B B e e d d L L o o c c a a t t o o r r after it has been installed and mapped, you must remap to the new room or location.
• Always apply the brakes when a patient is getting into or out of the product to avoid instability.
• Always apply the brakes when the patient is unattended.
• Do not apply the brakes to slow or stop the product while it is in motion.
• Always lock the siderails in the full up position with the sleep surface horizontal in the lowest position when you transport
the product with a patient.
• Always unplug the power cord before you transport the product.
• Always release the brakes before you transport the product. Do not transport the product with the brakes applied.
• Do not transport the product laterally after you apply the steer lock pedal. The product cannot swivel when you transport
with steer lock.
• Only use retractable traction or fracture frames. Failure to use a retractable frame may result in injury to the patient and/
or damage the equipment.
• Always set the siderail position to make sure that a patient is safely in the product.
• Always lock the operator control panel and patient control panel when the patient is unattended.
• Always keep the siderails outside of the oxygen tent.
• Always use only hospital-grade electric equipment consuming 10A or less with the auxiliary power outlet (option). The
use of standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment.
• Do not use the 110V auxiliary power outlet (option) for life sustaining equipment.
• Always lock the control panel when you leave the patient unattended.
• Always lock the patient control panel when the patient's condition requires extra safety measures.
• Do not use bed exit to replace patient monitoring protocol. Bed exit is intended only to aid in the detection of a patient
exiting the product.
• Do not use bed exit with patients who weigh less than 50 lb (23 kg).
• Do not use the scale system reading as a reference for medical treatment. The scale system assists only in the
monitoring of the patient's weight variation.
• Do not set the scale to zero or weigh the patient when a support surface therapy is active. Motion from the support
surface functions may adversely affect the scale system performance.