Télécharger Imprimer la page

Stryker S3 MedSurg Bed with StayPut Frame Mode D'emploi page 68

Masquer les pouces Voir aussi pour S3 MedSurg Bed with StayPut Frame:

Publicité

Les langues disponibles

Les langues disponibles

Conducted RF IEC 61000-
4-6
Radiated RF IEC 61000-4-3
N N o o t t e e - - At 80 MHz and 800 MHz, the higher frequency range applies.
N N o o t t e e - - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the 3005 S S 3 3 MedSurg bed is used exceeds the
applicable RF compliance level above, the 3005 S S 3 3 MedSurg bed should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the 3005
S S 3 3 MedSurg bed.
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
EN
64
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the 3005 S S 3 3
MedSurg bed, including
cables, than the
recommended separation
distance calculated from the
equation appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(1.2) (√ P )
80 MHz to 800 MHz
D=(2.3) (√ P )
800 MHz to 2.5 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey
a
, should be less than
the compliance level in each
frequency range
b
.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
3005-009-005 Rev AA.1

Publicité