Télécharger Imprimer la page

Stryker S3 MedSurg Bed with StayPut Frame Mode D'emploi page 60

Masquer les pouces Voir aussi pour S3 MedSurg Bed with StayPut Frame:

Publicité

Les langues disponibles

Les langues disponibles

C C l l e e a a n n i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the patient from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
• Always clean V V e e l l c c r r o o
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
To remove undesirable build-up before disinfecting between uses:
1. Hand wash all surfaces on the product with a mild detergent using spray or pre-soaked wipes.
2. Clean all exposed surfaces.
3. Follow the cleaning solution manufacturer's instructions for appropriate contact time and rinsing requirements.
4. Dry the product thoroughly before returning the product to service.
Avoid over saturation. Do not allow the product to remain wet.
EN
56
® ®
after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
3005-009-005 Rev AA.1

Publicité

loading

Ce manuel est également adapté pour:

3005