6.5.5. Programming the "END" ramp rate
To end a firing program, "END" must be programmed in the last segment.
Display
Pr. 1
Indicator in the
Heating ramp
150°C/HR
Cooling ramp
Heating ramp
END
Cooling ramp
h
g
*The keys
or
can also be held down for faster input.
Meaning
Description
Program
Select the firing program that is to
number
be changed.
Select the segment with the
segment
heating ramp to be changed into
display
a cooling ramp.
The ramp rate for the heating or
cooling ramp is shown in the
main display.
To mark the end of the program,
f
push the
appears in the main display.
key until "END"
67 / 317 - EN -
Remark
See Section 6.5.1.
See Section 6.5.1.
Possible main display:
„0,1 °C/h-9,9 °C/h",
„10 °C/h-999 °C/h"
• "END" ends the firing program.
• The firing program ends with the
segment in which "END" has been
programmed.
• After "END" has been
programmed, the soak/target
temperature or soak time can no
longer be selected.
• The "END" value is one step below
the ramp rate "0,1 °C/h".
Note:
If "END" is not entered in the last
segment, the controller issues the
error message "Error P" when the
program starts.