6.5.4. Programming the "FULL" ramp rate
An unregulated full load heating/cooling ramp ("FULL") can be programmed.
Display
Pr. 1
150°C/HR
FULL
h
g
*The keys
or
can also be held down for faster input.
Meaning
Description
Program
Select the firing program that is to be
number
changed.
Indicator in
Select the segment with the heating
the segment
ramp to be changed into a cooling
display
ramp.
Heating ramp
The ramp rate for the heating or
cooling ramp is shown in the main
Cooling ramp
display.
Heating ramp
Press or hold down the
"FULL" appears in the main display.
Cooling ramp
d
key until
66 / 317 - EN -
Remark
See Section 6.5.1.
See Section 6.5.1.
Possible main display:
„0,1 °C/h-9,9 °C/h",
„10 °C/h-999 °C/h"
• "FULL" heats or cools as fast as
possible.
• The "FULL" value is one step
above the ramp rate "999°C/h".