6 Maintenance and Care

6.1.- Waste water discharging
A Led is lighted on front panel when waste water level in tank reaches the allowable
maximum.
In this case the autoclave is inhibited.
Connect discharge tube to output connector (see fig. 2) and discharge water.
6.2.- Replacing the Sterile Filter
Replace Sterile Filter every three months, or in a shorter time if its colour is dark.
6.3.- Cleaning the Chamber
Clean the chamber with a slight abrasive cloth to remove solid particles that may be present.
Use only demineralised water to moisten the cloth.
6.4.- Door Regulation
To regulate the door lock pressure, it is necessary to adjust the regulator of the same door
using the double function wrench (extraction tray + wrench) provided.
Turn anti clockwise (fig.A) to increase lock pressure.
Turn clockwise (fig.B) to decrease the lock pressure.
FIG. A
FIG. B